Detector 3: Difference between revisions

From HelioWiki Home Page
Jump to navigation Jump to search
imported>Lglesener
No edit summary
imported>Lglesener
No edit summary
Line 5: Line 5:
===Notes===
===Notes===


Detector 3 is SEGMENTED following the 2014 anneal (yay!)
* 2014-Aug-26: D3 is suffering a lot from the known light leak problem.  This does not affect the quality of the data but greatly reduces livetime at times when light leaks in through the cryostat port.  To alleviate this problem, the G3 rear fast threshold was raised from 0x45 to 0x50 at 2014-238-19:22:56 UTC.


===D3 front slow LLD threshold history===
===D3 front slow LLD threshold history===

Revision as of 22:52, 4 September 2014

UPDATE IN PROGRESS

This page contains information about changes/issues for RHESSI detector 3 after the 2014 anneal. Detector updates from 2012-mid 2014 can be found here: Detector 3 anneal 2012

Notes

Detector 3 is SEGMENTED following the 2014 anneal (yay!)

  • 2014-Aug-26: D3 is suffering a lot from the known light leak problem. This does not affect the quality of the data but greatly reduces livetime at times when light leaks in through the cryostat port. To alleviate this problem, the G3 rear fast threshold was raised from 0x45 to 0x50 at 2014-238-19:22:56 UTC.

D3 front slow LLD threshold history

post 2014 anneal 0x

D3 front fast LLD threshold history

post 2014 anneal 0x

D3 rear slow LLD threshold history

post 2014 anneal 0x

D3 rear fast LLD threshold history

post 2014 anneal 0x
GAP IN THRESHOLD VALUES - ADD IN LATER
2014-08-29 18:33:28 UTC 0x60

D3 HV history

post 2014 anneal 3995 V