Revision as of 22:52, 4 September 2014 by imported>Lglesener
UPDATE IN PROGRESS
This page contains information about changes/issues for RHESSI detector 3 after the 2014 anneal. Detector updates from 2012-mid 2014 can be found here: Detector 3 anneal 2012
Notes
Detector 3 is SEGMENTED following the 2014 anneal (yay!)
- 2014-Aug-26: D3 is suffering a lot from the known light leak problem. This does not affect the quality of the data but greatly reduces livetime at times when light leaks in through the cryostat port. To alleviate this problem, the G3 rear fast threshold was raised from 0x45 to 0x50 at 2014-238-19:22:56 UTC.
D3 front slow LLD threshold history
D3 front fast LLD threshold history
D3 rear slow LLD threshold history
D3 rear fast LLD threshold history
| post 2014 anneal |
0x
|
| GAP IN THRESHOLD VALUES - ADD IN LATER
|
| 2014-08-29 18:33:28 UTC |
0x60
|
D3 HV history