Notes
Detector 6 had no notable issue since the 2012 anneal until the noise/livetime problem occurred in 2014 Jan.
On 2014 Jan 3, Hugh Hudson noticed that the attenuator system has been in state 3 for a few days. It turned out that, in Detector 6 (both front and rear), livetime was decreasing rapidly and preamp-reset and FAST LLD were also rapidly increasing (See also Tohban_Report_2014-01-08). Then, over the next few days, several adjustments were made to D6 high voltage and FAST threshold (both front and rear) as described below.
- 2014 Jan 04: As a first step, D6 HV was lowered by 8 steps down to about 3800 V.
14-004-06:54:40 /ihvdac detector=6, voltage=198
- 2014 Jan 05: The livetime kept decreasing rapidly and so the D6 HV was lowered again by another 20 steps down to 2671 V.
14-005-08:14:37 /ihvdac detector=6, voltage=148 (2868 V)
14-005-08:16:00 /ihvdac detector=6, voltage=138 (2671 V)
- 2014 Jan 06: Previous commandings led to >90% livetime but the FAST LLD was still somewhat high compared to that of D1. Because lowering HV further could result in faster degradation of the detector and also because David Smith reported that the HV changes worsened the spectroscopy (See his email on Jan 5), it was decided not to lower HV anymore. Instead, the front fast threshold was raised from 0x22 (~ 8 keV) to 0x50 (~ 20 keV).
14-006-08:03:32 /idpudumptabl dibtbl6
14-006-08:03:57 /idputable6 frontfastdac
14-006-08:04:08 /idpuload value=0x50
- 2014 Jan 08: The livetime has now reached ~98% and the FAST LLD reached its nominal level. However, it was noticed that the rear segment also had the same noise/livetime problem. Thus, rear fast threshold was raised from 0x45 (? keV) to 0x60 (? keV).
14-008-07:35:20 /idputable6 rearfastdac
14-008-07:35:40 /idpuload value=0x55
14-008-07:36:03 /idputable6 rearfastdac
14-008-07:36:56 /idpuload value=0x60
- 2014 Jan 08: In the meantime, it was decided to bring back the HV as much as possible to avoid further degradation of the detector. Over the course of 2 passes, the HV was raised by 30 steps up to 3259 V.
14-008-07:37:49 /ihvdac detector=6, voltage=148
14-008-07:39:15 /ihvdac detector=6, voltage=158 (3063 V)
14-008-09:15:22 /ihvdac detector=6, voltage=168 (3259 V)
- 2014 Jan 08: Because noise level seemed to have increased during the HV commandings, front fast threshold was briefly raised from 0x50 to 0x64 then back to 0x50.
14-008-09:18:13 /idputable6 frontfastdac
14-008-09:18:39 /idpuload value=0x64
14-008-09:24:57 /idputable6 frontfastdac
14-008-09:25:28 /idpuload value=0x50
- 2014 Jan 28: The D6 front fast LLD threshold was raised from 0x50 to 0x60 at 22:13 UTC.
- 2014 Jan 29: The D6 front fast LLD threshold was raised from 0x60 to 0x70 at 22:11 UTC.
- 2014 Feb 5: The D6 front slow LLD threshold was raised from 0x0C to 0x1A (~6 keV) at 20:27 UTC.
- 05-Mar-2014: D6 front fast LLD was raised 0x70 to 0x80 at 2014/058 19:25 UTC. In D6 rear fast valids, no commanding was done to this segments, but the rear fast valids have risen to to ~1,000 --> 5,000 counts over the week, which is now comparable to D9r fast rates. Their lifetime is still at 98%.
- 2014 Mar 10: D6 front fast threshold was raised from 0x80 to 0x90 at 2014-070-03:00 UTC due to rapidly decreased livetime (down to 10% at the time of the commanding). The livetime immediately recovered to ~80%.
- 2014 Mar 19: Several changes were made to D6 this week. The fast threshold was raised. This resulted in a surge in fast rates (150k!) and corresponding decrease in livetime. This is due to an issue seen before, i.e. that before the threshold raise there were so many rates that the system was essentially hung and not recording the true rate. Raising the threshold improved the situation, even though the consequences make it look like it's the opposite! A test was then performed whereby detector 6's HV was dropped to 2474V, left there for one orbit, and then raised again. In this orbit the fast rates dropped dramatically, freeing up livetime (changed from single digits up to ~97%), and reset rates dropped, though by <1 order of magnitude. The voltage was not brought all the way back up to its prior value; instead it was left at 3063V. The following day, the fast threshold was raised again because the livetime had slowly decayed again.
14-076-20:34:17 /IDPUTABLE6 FRONTFASTDAC 14-076-20:34:33 /IDPULOAD VALUE=0xA0 ;Was 0x90
14-076-23:45:02 /IHVDAC DETECTOR=6, VOLTAGE=148 (lowered from 168) 14-076-23:46:05 /IHVDAC DETECTOR=6, VOLTAGE=128 (2474 V)
14-077-01:23:14 /IHVDAC DETECTOR=6, VOLTAGE=138 (raised from 128) 14-077-01:24:40 /IHVDAC DETECTOR=6, VOLTAGE=148 14-077-01:26:08 /IHVDAC DETECTOR=6, VOLTAGE=158 (voltage ramp halted at 3064 V)
14-077-23:30:17 /IDPUTABLE6 FRONTFASTDAC 14-077-23:30:46 /IDPULOAD VALUE=0xB0
14-080-23:00:26 /IDPUTABLE6 FRONTFASTDAC
14-080-23:00:51 /IDPULOAD VALUE=0xC0
- 2014 Mar 27: Trying to reduce the large number of slow events
14-086-18:17:40 /IDPUTABLE6 FRONTSLOWDAC
14-086-18:17:52 /IDPULOAD VALUE=0x20
14-088-19:16:26 /IDPUTABLE6 FRONTSLOWDAC
14-088-19:16:40 /IDPULOAD VALUE=0x30
- 2014 Apr 7: HV decrease to reduce slow valid rates and fast valid rates
14-097-17:15:14 /IHVDAC DETECTOR=6, VOLTAGE=138 ; was 148
- 2014 Apr 10: Front fast threshold was raised to 0xD0
- 2014 Apr 11: Front fast threshold was raised again (this time to 0xE0) because previous change did not improve the situation.
- 2014 Apr 13: HV decrease by 10 counts (128) down to 2475V.
- 2014 Apr 15: Front fast threshold raised to 0xF0. The livetime recovered from 1 to 80%.
- 2014 Apr 18: Detector 6 HV was dropped by 20 counts to a value of 108 (~2100 V).
- 2014 May 17:
- G6 front fast threshold from 0xF0 to 0xFF (max)
- 14-137-04:03:49 /IDPUDUMPTABL TABLE=DIBTBL6
- 14-137-04:03:50 /IDPUTABLE6 OFFSET=FRONTFASTDAC
- 14-137-04:03:57 /IDPULOAD VALUE=0xFF
- 2014 May 19:
- An HV "experiment" was performed on D6; full details can be found here.
- 2014 May 22: G6 HV was decreased by 20 counts to a value of 88 ( ~1700V)
2014-142-21:39:28 /IHVDAC DETECTOR=6, VOLTAGE=88 ; was 108
- 2014 May 27: G6 Front slow threshold was changed to 0x3A
14-147-16:57:10 /IDPUTABLE6 OFFSET=FRONTSLOWDAC
14-147-16:57:26 /IDPULOAD VALUE=0x3A
- 2014 May 28: G6 HV decrease 10 counts to 78 (~1500 V)
- 14-148-21:43:21 /IHVDAC DETECTOR=6, VOLTAGE=78 ;(1495 V)
D6 front slow LLD threshold history
post 2012 anneal |
0x0C
|
2014-Feb-05 20:27 UTC |
0x1A (was 0x0C)
|
2014-Mar-27 18:17:52 UTC |
0x20 (was 0x1A)
|
2014-Mar-29 19:16:40 UTC |
0x30 (was 0x20)
|
2014-May-27 16:57:26 UTC |
0x3A (was 0x30)
|
D6 front fast LLD threshold history
post 2012 anneal |
0x22
|
2014-Jan-06 08:04:08 UTC |
0x50 (was 0x22)
|
2014-Jan-06 09:18:39 UTC |
0x64 (was 0x50)
|
2014-Jan-06 09:25:28 UTC |
0x50 (was 0x64)
|
2014-Jan-28 22:13 UTC |
0x60 (was 0x50)
|
2014-Jan-29 22:11 UTC |
0x70 (was 0x60)
|
2014-Feb-27 19:25 UTC |
0x80 (was 0x70)
|
2014-Mar-11 03:00 UTC |
0x90 (was 0x80)
|
2014-Mar-17 20:34:33 UTC |
0xA0 (was 0x90)
|
2014-Mar-18 23:30:17 UTC |
0xB0 (was 0xA0)
|
2014-Mar-21 23:00:51 UTC |
0xC0 (was 0xB0)
|
2014-Apr-10 18:09:44 UTC |
0xD0 (was 0xC0)
|
2014-Apr-11 17:55:11 UTC |
0xE0 (was 0xD0)
|
2014-Apr-15 16:55:00 UTC |
0xF0 (was 0xE0)
|
2014-May-17 04:03:25 UTC |
0xFF (was 0xF0)
|
D6 rear slow LLD threshold history
D6 rear fast LLD threshold history
post 2012 anneal |
0x45
|
2014-Jan-08 07:35:40 UTC |
0x55 (was 0x45)
|
2014-Jan-08 07:36:56 UTC |
0x60 (was 0x55)
|